Back to top
Sensocon USA Series  DPT1-R8 - Range -0.1 - 0.1 in WC
Sensocon USA Series  DPT1-R8 - Range -0.1 - 0.1 in WC

Sensocon USA Series DPT1-R8 - Range -0.1 - 0.1 in WC

Sensocon USA Series DPT1-R8 - Range -0.1 - 0.1 in WC Specification

  • Grade
  • Industrial Grade
  • Outer Size
  • 120 x 90 x 36 mm (approx.)
  • Seal Material
  • Silicone
  • Thread
  • 1/8 NPT Female Pressure Connections
  • Display
  • LCD Digital Display
  • Sensor Type
  • Differential Pressure Sensor
  • Material
  • ABS Plastic Case
  • Processing Type
  • Electronic
  • Mount Type
  • Surface/Panel Mount
  • Scale Range
  • -0.1 to 0.1 in WC
  • Accuracy
  • 1% of Full Scale
  • Finishing
  • Matte Grey Housing
  • Measure Pressure Of
  • Air and Non-Corrosive Gases
  • Accessories
  • Mounting Screws, Manual
 

Sensocon USA Series DPT1-R8 - Range -0.1 - 0.1 in WC Trade Information

  • Payment Terms
  • Cash on Delivery (COD), Cash Advance (CA), Cash Against Delivery (CAD), Cash in Advance (CID)
  • Main Export Market(s)
  • Asia
  • Main Domestic Market
  • Dadra and Nagar Haveli, Himachal Pradesh, Meghalaya, Andaman and Nicobar Islands, Nagaland, South India, North India, East India, West India, Andhra Pradesh, Assam, Arunachal Pradesh, Bihar, Chandigarh, Goa, Jammu and Kashmir, Jharkhand, Karnataka, Mizoram, Manipur, Punjab, Pondicherry, Rajasthan, Sikkim, Tamil Nadu, Tripura, West Bengal, Uttarakhand, Daman and Diu, Lakshadweep, Uttar Pradesh, Maharashtra, Haryana, Delhi, Gujarat, Madhya Pradesh, Telangana, Kerala, Central India, Odisha, Chhattisgarh, All India
 

About Sensocon USA Series DPT1-R8 - Range -0.1 - 0.1 in WC

Sensocon USA Series  DPT1-R8 - Range -0.1 - 0.1 in WC 

APPLICATIONS 

DPT-R8 series devices are commonly used in HVAC/R systems for: Fan, blower and filter monitoring Pressure and flow monitoring Valve and damper control Pressure monitoring in clean rooms

WARNING 

READ THESE INSTRUCTIONS CAREFULLY BEFORE ATTEMPTING TO INSTALL, OPERATE OR SERVICE THIS DEVICE.

Failure to observe safety information and comply with instructions could result in PERSONAL INJURY, DEATH AND/ OR PROPERTY DAMAGE.

To avoid electrical shock or damage to equipment, disconnect power before installing or servicing and use only wiring with insulation rated for full device operating voltage.

To avoid potential fire and/or explosion do not use in potentially flammable or explosive atmospheres.

Retain these instructions for future reference.

This product, when installed, will be part of an engineered system whose specifications and performance characteristics are not designed or controlled by Sensocon. Review applications and national and local codes to assure that the installation will be functional and safe. Use only experienced, knowledgeable, technicians to install this device. 

SPECIFICATIONS 

Performance Accuracy (at applied pressure) Ranges < 0.5 inWC = 0.01 inWC Ranges 0.5 inWC = 1.5% inWC Accuracy specifications include: general accuracy, temperature drift, linearity, hysteresis, long term stability, and repetition error

Thermal Effects Temperature compensated across the full spectrum of capability 

 Overpressure Proof Pressure: 100 inWC Burst Pressure: 120 inWC

Zero Point Calibration Automatic with function or Manual push-button Auto-Zero

Response Time 4.0 0.8 sec, Selectable via Jumper

Technical Specifications Media Compatibility Dry air or non-aggressive gases

Measuring Units (select via Jumper) Pa, kPa, mbar, inWC, mmWC Measuring Element Piezoresistive

Environment: Operating Temperature: 14 122F (-10 50C) Storage Temperature: -4 158F (-20 70C) Humidity: 0 to 95 % rH, non condensing 

Physical Case Dimensions: 3.54 x 3.74 x 1.4 (90.0 x 95.0 x 36.0 mm) Weight: 5.3 oz (150 g) Mounting: 2 each 0.1875 (4.3 mm) screw holes, one slotted Materials: Case: ABS (UL 94 V-0 Approved) Lid: PC (UL 94 V-1 Approved) Protection standard: IP54 / NEMA3 

Display (optional) 3 1/2 Digit LCD Backlit Display 2-Line Display (12 character/line) Size: 1.81 W x 0.57 H (46.0 x 14.5 mm) 

Electrical connections: 4 screw terminal block Wire: 12-24AWG (0.2 1.5 mm2 

Cable Entry Strain Relief: M16 Knockout : 0.625 (16mm)

Pressure Fittings 3/16 (5.2mm) barbed brass + High pressure Low pressure 

Electrical Voltage Circuit: 3-wire ([V Out], 24V, GND) Input: 24VAC or VDC, 10% Output: 0-10V or 0-5V, selectable via Jumper Power Consumption: <1.0W Resistance Minimum: 1k

Current Circuit: 3-wire (mA Out, 24V, GND) Input: 24VAC or VDC, 10% Output: 4-20 mA Power Consumption: <1.2W Maximum load: 500 



Highly Precise Differential Pressure Measurement

The DPT1-R8 delivers exceptional accuracy and a fine resolution, enabling users to detect even minor pressure changes. Its rapid response time and robust digital display ensure reliable real-time feedback, making it suitable for sensitive environments such as clean rooms and HVAC systems.


Flexible Output and Easy Configuration

Equipped with selectable 4-20 mA or 0-10 V outputs, the sensor adapts effortlessly to diverse monitoring setups. Digital zero and span adjustment buttons make calibration straightforward, significantly reducing setup times and maintenance efforts.


Built for Durability and Versatility

Housed in a matte grey ABS plastic case, the DPT1-R8 is designed for industrial settings. It features an IP54 rating for the front panel, enhancing protection against dust and splashes, and supports mounting on both panels and surfaces for versatile installation.

FAQs of Sensocon USA Series DPT1-R8 - Range -0.1 - 0.1 in WC:


Q: How do I install the Sensocon USA DPT1-R8 differential pressure sensor?

A: The DPT1-R8 can be installed on either a surface or a panel using the included mounting screws. Simply connect the 1/8 NPT female pressure ports to your air or non-corrosive gas lines, and wire the power supply (24V AC/DC) and output signal according to the manual.

Q: What applications is the DPT1-R8 sensor best suited for?

A: This sensor is ideal for monitoring and controlling pressure in HVAC systems, clean rooms, filter monitoring, and building automation. Its industrial-grade accuracy makes it especially useful where precise differential pressure monitoring is critical.

Q: When should I use the zero/span adjustment function?

A: Use the digital push-button zero/span adjustment during setup, maintenance, or when you notice measurement drift. This ensures the sensor maintains accurate and reliable readings over time.

Q: Where can I mount the DPT1-R8 in my facility?

A: You can mount the DPT1-R8 on both flat surfaces and control panels, providing flexibility for different installations within your facility. Its compact size and durable housing allow for placement in various industrial or commercial environments.

Q: What is the benefit of having selectable 4-20 mA and 0-10 V output signals?

A: Selectable output signals give you compatibility with a wide range of industrial automation systems and controllers, allowing easy integration regardless of your systems requirements.

Q: How can I ensure optimal performance and longevity of the sensor?

A: To maintain optimal performance, install the sensor away from sources of vibration and moisture, ensure connections are tight, and periodically check calibration using the digital adjustment features. Clean with a dry cloth as needed to keep the LCD display visible.

Tell us about your requirement
product

Price:

Quantity
Select Unit

  • 50
  • 100
  • 200
  • 250
  • 500
  • 1000+
Additional detail
Mobile number

Email